COD, TOC determination using UV spectrophotometry and its application to monitor the potassium hydrogen phthalate degradation by H2 O2 -HCO3 – oxidation system

Authors

  • Thi Bich Viet Nguyen*, Bich Ngan Nguyen, Thi Theu Tran
  • Thi Diu Vu

Keywords:

COD, H2 O2 -HCO3 – oxidation system, KHP, TOC, UV spectrophotometry

Abstract

In this study, the COD and TOC in H2O2-HCO3 oxidation system containing potassium hydrogen phthalate were determined by UV spectrophotometry (260-310 nm). The pH and H2O2 concentrations were investigated as factors influencing the absorbance measurements. The obtained standard curves were Abs = (3.10±0.04)x103xCOD - (0.015±0.003) (R2=0.9996) with LOD of 5.1 mg O2/l and LOQ of 13.6 mg O2/l, and Abs = (0.008±0.0001)xTOC - (0.015±0.003) (R2=0.9996) with LOD of 1.6 mg/l and LOQ of 5.4 mg/l. The method was applied to monitor the degradation of potassium hydrogen phthalate by the H2O2-HCO3 oxidation system. The results revealed that the COD and TOC removal efficiencies reached ~85% after 90 minutes of UVC irradiation. The UV spectra and HPLC chromatographs showed that no aromatic compounds were obtained in the degradation products.

Downloads

Download data is not yet available.

Author Biographies

  • Thi Bich Viet Nguyen*, Bich Ngan Nguyen, Thi Theu Tran

    Faculty of Chemistry, Hanoi National University of Education

  • Thi Diu Vu

    Faculty of Science, Thai Binh College of Education and Training

Published

2021-12-03