Research and development of 40 GHz bandwidth laser relative intensity noise measurement system using PID controller and ESA spectrum analyser

Authors

  • Hai Long Ngo*, Hoang Dat Luu, Dinh Chi Tran, Viet Hoang Do, Van Binh Le, Khac Thien Cao, Quoc Tuan Banh

Keywords:

ESA, laser, PID controller, relative intensity noise.

Abstract

Measurement system is essential. Relative intensity noise (RIN) is a parameter that directly affects the system quality. Therefore, RIN needs to be measured
periodically. This study presents a high-sensitivity, broadband (40 GHz) RIN measurement system using a precisely temperature-controlled photodetector and
an electrical spectrum signal analyser...